Michigan B184 Update June 18, 2004 |
Back to B184 updates * Previous * Next
1) Chamber passing CR test
a) Chamber passed CR test this week: none
b) Total number of chamber passed CR test at CERN: 26
2) chamber in progress
EMS4C06: Retaking CR data
EML3A13: CR data has been taken and analyzed, redo the mezz scan
with HV off and on.
EMS5C06: Ready for CR test
EML3A11: Ready for CR test
EML3A05: T sensors and inplane system tested, dark current measured.
3) Chambers in hospital
EMS5A10: After epoxy cured for three days, we pressurized it with
helium to 3 bar. Vecco was used to scan the leaks. Leak rate
was reduced by half. Continue to debug the leaks.
4) Problems and other issues
a) Tubes having high dark current
EMS5C06 and EML3A13 have 12 and 4 tubes with high dark current
respectively. The dark current ranges from 100 na to over 1000
na. All are being treated with -HV=3.2-3.4 kV.
It's extremely time cosuming to deal with chambers with so many
tubes having high dark current.
b) Parts and tools
- since we have preassembled all FC and motherboards, we are able
to count our parts once more. This time should be more precise.
A list of parts UM requested has been sent to Bing.
- We are running out of SHH boards. A message has been sent to
Robert.
- We do find another box of B sensor platform.
- Since UW is sharing all our parts, parts inventory will be more
difficult. We have a notebook to register all they take.
- A HV power supply was broken by UW.
- A Q196 is broken by UW.
c) There were twice the power cords were unplugged without
notifying the users. This cause lost our time for CR test.
A sign is now put on the switch board please "don't disconnect the
power cord "
We now decide to the fix extension cord to each station.
This is week we fight toubles, leaks, high dark current, rejecting
mezz cards, .....
d) Reminders
- Don't tune any devices before consulting experts. The low
voltage power supply on DAQ system for CR test is not allowed to
be tuned.
- Please inform me before you bring any tools and parts away from
bldg 184. This is just to avoid spending lots of time searching
around.
- As you can see, if we don't work out the chambers for CR
test. We don't have much to do, so speeding up the CR test is
crucial again.
Zhengguo Zhao